Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8024B: (2 year) CAL VER /5520 DATE: 08-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 56 NUMBER OF LINES: 215 CONFIGURATION: Fluke 5520A CONFIGURATION: Fluke 45 ============================================================================= # # Source: # Fluke 8024B instruction manual (PN 616052 July 1981). # # Compatibility: # 5520/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Turn the UUT on. 1.006 DISP Ambient temperature: 21C - 25C. 1.006 DISP Relative humidity: less than 80%. 1.006 DISP Allow the UUT to stabilize. 1.007 HEAD {DC VOLTAGE TESTS} 1.008 DISP Connect the 5520A and the UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT V/Ohm/S 1.008 DISP [32] 5520A NORMAL LO to UUT COMMON 1.009 HEAD DC VOLTAGE TESTS: {200mV Range} 1.010 5520 200 190.0mV 0.3U 2W 2.001 5520 200 -190.0mV 0.3U 2W 3.001 HEAD DC VOLTAGE TESTS: {2V Range} 3.002 5520 2 0.000V 0.001U 2W 4.001 5520 2 1.900V 0.003U 2W 5.001 HEAD DC VOLTAGE TESTS: {20V Range} 5.002 5520 20 19.00V 0.03U 2W 6.001 HEAD DC VOLTAGE TESTS: {200V Range} 6.002 5520 200 190.0V 0.3U 2W 7.001 HEAD DC VOLTAGE TESTS: {1000V Range} 7.002 5520 1000 1000V 2U 2W 8.001 HEAD {AC VOLTAGE TESTS} 8.002 HEAD AC VOLTAGE TESTS: {200mV Range} 8.003 RNG 200mV AC 8.004 5520 0.0mV S 2W 8.005 MEMI Enter UUT reading in millivolts AC: 8.006 MEME 8.007 MEMC mV +0.2U 9.001 5520 200 190.0mV 1.6U 100H SI 2W 10.001 5520 200 190.0mV 3.2U 2kH SI 2W 11.001 5520 200 190.0mV 9.9U 5kH SI 2W 12.001 HEAD AC VOLTAGE TESTS: {2V Range} 12.002 5520 2 1.900V 0.099U 5kH SI 2W 13.001 5520 2 1.900V 0.032U 2kH SI 2W 14.001 5520 2 1.900V 0.016U 100H SI 2W 15.001 5520 2 0.190V 0.015U 5kH SI 2W 16.001 HEAD AC VOLTAGE TESTS: {20V Range} 16.002 5520 20 19.00V 0.16U 100H SI 2W 17.001 5520 20 19.00V 0.32U 2kH SI 2W 18.001 5520 20 19.00V 0.99U 5kH SI 2W 19.001 HEAD AC VOLTAGE TESTS: {200V Range} 19.002 5520 200 100.0V 1.8U 2kH SI 2W 20.001 5520 200 100.0V 1U 100H SI 2W 21.001 HEAD AC VOLTAGE TESTS: {750V Range} 21.002 5520 750 750V 10U 100H SI 2W 22.001 5520 750 750V 10U 1kH SI 2W 23.001 HEAD {DC CURRENT TESTS} 23.002 DISP Connect the 5520A and the UUT as follows: 23.002 DISP [32] 5520A AUX HI to UUT mA 23.002 DISP [32] 5520A AUX LO to UUT COMMON 23.003 HEAD DC CURRENT TESTS: {2mA Range} 23.004 5520 2 1.900mA 0.015U 2W 24.001 HEAD DC CURRENT TESTS: {20mA Range} 24.002 5520 20 19.00mA 0.15U 2W 25.001 HEAD DC CURRENT TESTS: {200mA Range} 25.002 5520 200 190.0mA 1.5U 2W 26.001 5520 200 -190.0mA 1.5U 2W 27.001 HEAD DC CURRENT TESTS: {2000mA Range} 27.002 5520 2000 1900mA 15U 2W 28.001 HEAD {AC CURRENT TESTS} 28.002 HEAD AC CURRENT TESTS: {2mA Range} 28.003 RNG 2mA AC 28.004 5520 0.000mA S 2W 28.005 MEMI Enter UUT reading in milliamps AC: 28.006 MEME 28.007 MEMC mA +0.002U 29.001 5520 2 1.900mA 0.059U 400H SI 2W 30.001 5520 2 1.900mA 0.059U 100H SI 2W 31.001 HEAD AC CURRENT TESTS: {20mA Range} 31.002 5520 20 19.00mA 0.31U 100H SI 2W 32.001 5520 20 19.00mA 0.31U 1kH SI 2W 33.001 HEAD AC CURRENT TESTS: {200mA Range} 33.002 5520 200 190.0mA 3.1U 1kH SI 2W 34.001 5520 200 190.0mA 3.1U 100H SI 2W 35.001 HEAD AC CURRENT TESTS: {2000mA Range} 35.002 5520 2000 1900mA 31U 100H SI 2W 36.001 5520 2000 1900mA 31U 1kH SI 2W 37.001 HEAD {RESISTANCE TESTS} 37.002 DISP Connect the 5520A and the UUT as follows: 37.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/S 37.002 DISP [32] 5520A NORMAL LO to UUT COMMON 37.002 DISP [32] 5520A AUX HI to UUT V/Ohm/S 37.002 DISP [32] 5520A AUX LO to UUT COMMON 37.003 HEAD RESISTANCE TESTS: {200 Ohm Range} 37.004 5520 200 100.0Z 0.5U CW 38.001 HEAD RESISTANCE TESTS: {2 kOhm Range} 38.002 5520 2 1.000kZ 0.002U CW 39.001 HEAD RESISTANCE TESTS: {20 kOhm Range} 39.002 5520 20 10.00kZ 0.02U CW 40.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 40.002 5520 200 100.0kZ 0.2U CW 41.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 41.002 DISP Connect the 5520A and the UUT as follows: 41.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/S 41.002 DISP [32] 5520A NORMAL LO to UUT COMMON 41.003 5520 2000 1000kZ 3U 2W 42.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 42.002 5520 20 10.00MZ 0.21U 2W 43.001 HEAD {CONDUCTANCE TEST} 43.002 HEAD CONDUCTANCE TEST: {200nS Range} 43.003 5520 200 100.0nY 3U 2W 44.001 HEAD {PEAK HOLD TEST} 44.002 RNG 2V AC 44.003 5520 1.900V 100H SI S 2W 44.004 DISP Push the Peak Hold switch ON. 44.005 MATH MEM1 = 1.900 44.006 MEMI Enter UUT reading in volts: 44.007 MEME 44.008 MEMC V 0.067U 45.001 5520 1.0mV 100H SI S 2W 45.002 DISP Count the number of digits the UUT changes 45.002 DISP prior to the beeper sounding.[D10000][7] 45.003 EVAL Did the UUT change less than 10 counts? 46.001 DISP Push the Peak Hold switch OFF. 46.002 HEAD {CONTINUITY TEST AUDIBLE TONE} 46.003 RNG 2kZ 46.004 EVAL Is the " ^ " indicator ON? 47.001 5520 0Z S CW 47.002 EVAL Is the " ^ " indicator OFF and the " v " indicator ON? 48.001 DISP Depress the AC/DC switch to 48.001 DISP activate the audible tone. 48.002 5520 0Z S CW 48.003 EVAL Is there an audible tone? 49.001 HEAD {BT INDICATOR} 49.002 DISP Connect a variable DC power supply to the Battery 49.002 DISP Elminator Connector. 49.002 DISP Connect the Fluke 45 to the output of the variable 49.002 DISP power supply. 49.003 RNG 2kZ 49.004 5520 0Z S CW 49.005 DISP Adjust the variable power supply for an 45 reading of 49.005 DISP +10V. 49.006 45 10.0V S 2W 49.007 5520 1.000kZ 0.002U CW 50.001 DISP Adjust the variable power supply for an 45 reading of 50.001 DISP +6V. 50.002 45 6.0V S 2W 50.003 5520 1.000kZ 0.002U CW 51.001 DISP Connect the 5520A and the UUT as follows: 51.001 DISP [32] 5520A NORMAL HI to UUT V/Ohm/S 51.001 DISP [32] 5520A NORMAL LO to UUT COMMON 51.002 RNG 200mV 51.003 5520 190.0mV S 2W 51.004 DISP Adjust the variable power supply until the BT indictor 51.004 DISP appears in the UUT display. 51.005 45 7.0V 0.5U 2W 52.001 5520 190.0mV S 2W 52.002 MESS Decrease the output of the variable supply until the 52.002 MESS 8842A displays +6V. 52.003 45 6.0V S 2W 52.004 5520 190.0mV 0.2U 2W 53.001 HEAD {LEVEL DETECTOR TEST} 53.002 RNG 200kZ 53.003 DISP Remove all connections to the 8024B and insert 53.003 DISP a test lead set into the V Ohm S and Common input 53.003 DISP terminals. As quickly as possible, strike the 53.003 DISP tips of the test leads together. Verify that the 53.003 DISP " v " indicator blinks each time the probe tips 53.003 DISP touch. 53.004 EVAL Did the " v " indicator blink each time? 54.001 DISP Repeat the test holding the test leads together 54.001 DISP and momentarily separating them. Insure that the 54.001 DISP " ^ " indicator blinks each time the connection 54.001 DISP is removed. 54.002 EVAL Did the " ^ " indicator blink each time? 55.001 HEAD {TEMPERATURE} 55.002 DISP Depress the TEMP degC range switch and release the 55.002 DISP kOhm/nS switch on the UUT. 55.003 DISP Connect the 80T-150 thermocouple accessory to the 55.003 DISP V/kOhm/nS and COMMON Inputs. 55.004 DISP Wipe the tip clean and place the tip between your thumb 55.004 DISP and index finger until the display readings stabilize. 55.005 EVAL Does the UUT read between 34degC and 39degC? 56.001 END